Region

Applications

EPA 552.1 by Strata-X-A on a ZB-XLB-HT

Application Detail (App ID: 19322)

Part No:7HM-G024-11
Phase:Proprietary XLB Phase
Dimensions:30 meters x 0.32 mm x 0.25 µm
Oven Profile:
Carrier Gas:Constant Flow Helium, 2.3 mL/min
Injection:Pulsed 0.0000000000:1 0 µL @ 250°C
Detection: Refractive Index (340°C)
Note:Derivatization: Add 1mL of MTBE to the eluent in a screw cap vial or test tube. Heat the sample at 50ºC for 1.5 hours. Remove and allow the sample to cool to room temperature. Slowly add 10mL of saturated sodium bicarbonate and invert the sample several times releasing carbon dioxide gas frequently. Aspirate 100µL of the organic layer and transfer to a 2mL vial with insert. Add 10µL of internal standard (1,2,3-Trichloropropane) and inject into GC-ECD system.


Analyte Details

Analyte Detail not available

1
Chloroacetic acid
search

2
Bromoacetic acid
search

3
Dichloroacetic acid
search

4
Dalapon
search

5
Trichloroacetic acid
search

6
Bromochloroacetic acid
search

7
1,2,3-Trichloropropane
search

8
Dibromoacetic acid
search